2 QSR Violations at Qinjiang Kingphar Medical

Share

A 6/30-7/3 FDA inspection at Qinjiang Kingphar Medical Material Co. in Qinjiang, China, identified two Quality System Regulation violations in the firm’s manufacturing of surgical masks, wound dressings, sterilization wraps, and surgical drapes. A 10/3 Warning Letter says the specific violations were:

  • failure to establish and maintain procedures to prevent contamination of equipment or products by substances that could reasonably be expected to have an adverse effect on product quality (repeat observation); and
  • failure to establish and maintain procedures to adequately control environmental conditions which could reasonably be expected to have an adverse effect on product quality, and to periodically inspect the environmental control system to verify that the system, including necessary equipment, is adequate and functioning properly.

Qinjiang was told to respond with a list of specific actions it has taken to correct the violations and prevent their recurrence, with documentation of each action and a schedule for completing the corrections.