Becton, Dickinson Recalls ChloraPrep Applicators
Becton, Dickinson (BD) has recalled its ChloraPrep 3 mL applicator due to possible fungal contamination, which only affects climate zone IV regions in specific U.S. territories (Puerto Rico, Guam, U.S. Virgin Islands, Northern Mariana Islands and American Samoa). These warm, high-humidity locations promote the growth of Aspergillus penicillioides, a recall notice says. The company notes that the recall does not include 3 mL applicators found in kits. ChloraPrep is a patient preoperative skin preparation.
BD says internal product quality testing identified that storage of the applicators “where product may be consistently exposed to temperatures of 30 degrees Celsius (86 degrees Fahrenheit) and 75% relative humidity for more than six months can result in the growth of Aspergillus penicillioides, a type of fungus, resulting in a breach in the outer package integrity,” the notice says. “Contamination of skin preparation products with Aspergillus penicillioides may lead to serious systemic infection, sepsis, illness and death.”