# Cook Tracheostomy Introducer Recall is Class 1
- URL: https://www.fdaweb.com/cook-tracheostomy-introducer-recall-is-class-1/
- Published: 2023-05-11T12:00:00.000Z
- Updated: 2026-09-14T18:31:32.000Z
- Author: David McFarland
- Tags: Devices, #legacy-id-D5154421

FDA says the Cook Medical 4/7 recall of the Blue Rhino G2Multi Percutaneous Tracheostomy Introducer is Class 1\. A [notice](https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=199620&ref=fdaweb.com) says the company recalled 3,053 units following customer reports that the tracheostomy in the sets and trays was not making a secure connection with the 15mm cap and other 15mm circuit components and accessories. “This resulted in an unsecure connection between the device connector and circuit components, cap, or accessories,” FDA says.