# Dexcom Glucose Monitor Software Correction is Class 1
- URL: https://www.fdaweb.com/dexcom-glucose-monitor-software-correction-is-class-1/
- Published: 2026-01-06T12:00:00.000Z
- Updated: 2026-09-14T15:29:53.000Z
- Author: David McFarland
- Tags: Devices, #legacy-id-D5160438

FDA says the Dexcom 9/18/2025 software correction for its Dexcom G6 and G6 Pro continuous glucose monitoring software is Class 1\. An agency [notice](https://www.fda.gov/medical-devices/medical-device-recalls-and-early-alerts/continuous-glucose-monitoring-software-correction-dexcom-issues-correction-dexcom-g6-and-g6-pro?ref=fdaweb.com) says the notice only affects users of version 1.15.0 of the Android U.S. app. The notice says Dexcom made the correction due to an identified bug that can cause the app to terminate unexpectedly.

“As a result,” the notice says, “the user may not receive estimated glucose values, alarms, or alerts. If a user is unaware that the app has terminated, there is potential for missed detection of a hyperglycemic or hypoglycemic event.”

As of 9/16, the company had not reported any serious injuries or deaths associated with the issue.