Draeger Anesthesia Workstation Correction Expanded
Draeger has expanded a 10/2024 device correction (Class 1 recall) to include additional Atlan A350 and A350XL anesthesia workstations used in operating rooms and other procedural settings. The devices are used to deliver inhaled anesthesia and provide both automatic and manual ventilation for adult, pediatric, and neonatal patients.
According to an FDA notice, the affected devices contain a manufacturing defect that may cause the piston ventilator to fail either before use or during a procedure. If the failure occurs while the device is in standby mode or undergoing system testing, mechanical ventilation may not start. If the malfunction occurs during active use, the workstation displays a "Ventilator error!!!" message and mechanical ventilation may stop.
Although ventilation support can continue through manual ventilation or spontaneous breathing modes, clinicians may need to manually ventilate patients to prevent injury, the agency says. It warns that ventilation failure could result in serious adverse health consequences, including hypoxia, loss of lung recruitment, bradycardia, cardiac arrest, and death.
Last month, Draeger notified customers of the issue and advised facilities that the devices may continue to be used under constant supervision until corrective actions are completed. The company plans to replace the ventilator motor assembly in affected anesthesia workstations.