Ethicon Endo-Surgery Recalls Surgical Stapler

Johnson & Johnson’s Ethicon Endo-Surgery has recalled (Class 1) its Endopath Echelon Vascular White Reload for Advanced Placement Tip surgical stapler due to increased reports regarding inadvertent instrument lockout during surgical procedures involving the device. “If lockout occurs, the device will momentarily activate but will not cut or staple tissue, and additional steps will be required to open it and remove it from tissue,” a recall notice says. “If user is unable to remove the device from the tissue, potential harms to patient include surgical delay, bleeding/hemorrhage, life-threatening hemorrhage/hemorrhagic shock, and conversion to open therapy.” The notice says two adverse events have been received, including one patient death.

Read more