FDA Guide on Transfer of Medical Device 510(k) Clearances
FDA has issued a final guidance clarifying when manufacturers and other companies acquiring a medical device's 510(k) clearance must submit a new premarket notification and update agency registration and listing information.
The document, Transfer of a Premarket Notification (510(k)) Clearance — Questions and Answers, explains that a company acquiring an existing 510(k) clearance generally does not need to submit a new 510(k) if the device has not been significantly changed in design, components, manufacturing methods or intended use.
However, the new clearance holder must meet applicable establishment registration and device listing requirements, including use of the original FDA-assigned 510(k) number. Significant device modifications generally require a new submission.
The guidance also notes that FDA recognizes only one 510(k) holder for a device at a time, although other companies, including contract manufacturers and relabelers, may use the clearance number when listing the device.
FDA further addresses labeling changes, unique device identification requirements and the responsibilities of companies involved in clearance transfers.