FDA Releases Magellan Diagnostics FDA-483
FDA has released the FDA-483 issued after a 5/10-6/29 inspection at Magellan Diagnostics, North Billerica, MA. The company was the subject of a 5/17 agency warning that Magellan Diagnostics LeadCare tests systems performed on blood drawn from a vein may provide inaccurate results.
An agency 7/13 statement (contains link to FDA-483) says the agency takes the inspection observations and risks the tests may have posed to patients very seriously. “We are carefully reviewing the evidence collected during the inspection to determine if there have been violations of federal law and whether further action is warranted,” the statement says.
Inspection observations on the FDA-483 were:
- design validation did not ensure that the device conforms to defined user needs and intended uses;
- risk analysis is inadequate;
- procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established;
- procedures for corrective and preventive action have not been adequately established;
- a report of the required information about device correction and removal actions was not sent to FDA within 10 days of the correction or removal;
- procedures for design changes have not been adequately established;
- acceptance criteria were not established before performing validation activities;
- a Medical Device Reporting report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur; and
- procedures have not been established to control product that does not conform to specified requirements.