FDA Warns of Embolism Risk From Heraeus Vascular Guidewires

Share

FDA has issued an early alert concerning a potentially high-risk problem with certain vascular guidewires manufactured by Heraeus Medical Components after the company identified a coating defect that could release particles into patients' bloodstreams and cause thromboembolic events.

The company issued voluntary recall letters 11/4–5 recommending that affected products be removed from use and distribution. As of 11/5, Heraeus had not reported any serious injuries or deaths associated with the issue, according to the alert.

The affected products include certain lots of Navilyst Vascular Guidewires, Spectrum Vascular Guidewires and Arrow Marked Guidewires.

According to FDA, the hydrophilic coating on the guidewires may swell inconsistently when hydrated with saline, causing the coating to clump, particularly around the distal coil section. The defect increases the potential for coating material to flake, become displaced or detach, releasing particles into the bloodstream.

The alert says the affected guidewires are used for percutaneous access to peripheral blood vessels using the Seldinger technique. They are not indicated for coronary or cerebral vascular procedures.