# GE Recalls Carestation Anesthesia Systems
- URL: https://www.fdaweb.com/ge-recalls-carestation-anesthesia-systems/
- Published: 2020-02-07T12:00:00.000Z
- Updated: 2026-09-14T16:02:13.000Z
- Author: David McFarland
- Tags: Devices, #legacy-id-D5146044

GE Healthcare has recalled the Carestation 600 Series Anesthesia Systems because there is a potential for a loose cable connection inside the system which may cause the mechanical ventilation to stop operating. If this happens, the system will sound a high priority audio and visual alarm to alert health care providers, a [recall notice ](http://s2027422842.t.en25.com/e/es?s=2027422842&e=300259&elqTrackId=376c7bc788024cd5a73d955f2e3dcbdc&elq=f3cea0bc5bd548efa3447f3c0b29818c&elqaid=11165&elqat=1&ref=fdaweb.com)says. Loss of mechanical ventilation could lead to low blood oxygen levels if the health care provider does not ventilate the patient manually or with another system, it says. To date, no injuries or deaths have been reported.