GE Recalls Ventilator Air Compressors

Share

GE Healthcare’s Datex-Ohmeda unit has recalled (Class 1) its EVair 03 (Jun-air) air compressor (model numbers: 1609000 and 1609002) after the company says it became aware of elevated levels of formaldehyde when the EVair or EVair 03 (Jun-Air) optional compressors are used with the CARESCAPE R860 or Engstrom Carestation/Pro ventilators, respectively.

“These elevated results were observed in preliminary testing that was conducted at an elevated room temperature of 40 degrees C (104 degrees F), at the lowest possible flow condition of 2 L/min (worstcase, minimum bias flow with no additional ventilation), and all of the gas being supplied from the compressor (i.e., FiO2 of 21% / no supplemental oxygen),” a recall notice says.

GE HealthCare says it is continuing to evaluate the root cause of the elevated levels. “While these test conditions are not representative of typical clinical use conditions, GE HealthCare is taking this action to further reduce the potential for patient exposure,” it adds.