Hamilton Recalls Defective Coaxial Breathing Sets
Hamilton Medical is recalling (Class 1) certain lots of coaxial breathing sets due to a quality issue involving a defective blade in the tube-cutting stage of production that may lead to a crack in the inner blue limb of the circuit. “The inner blue limb delivers fresh gas to the patient while the outer limb carries away exhaled gas,” a recall notice says, adding that if a leakage occurs from a crack in this limb, then separation between inspiratory and expiratory flows becomes compromised. This could lead to the partial or complete rebreathing of exhaled gases, the notice says.
Using affected products may cause serious adverse health consequences, including acute buildup of carbon dioxide (hypercapnia) that the body can’t remove (respiratory acidosis), according to the notice. “If the problem goes unrecognized, it may cause organ dysfunction and death,” it says. To date, there have been no reported injuries or deaths related to this issue.
The breathing circuits are intended to connect Hamilton-C1/T1/MR1 ventilators to a patient tube or respiratory mask during ventilation.