LivaNova Again Recalls LifeSPARC Controller Over Software

Share

LivaNova (TandemLife) has again recalled its LifeSPARC Controller, part of the LifeSPARC extracorporeal system used in heart surgeries, due to a problematic software update that was designed to address a previous software malfunction (see earlier story). The software update was found to create the same problem where the device erroneously detects frozen or unresponsive software and triggers a “Critical Failure” mode, which clears the controller screen and sets an alarm that cannot be muted or turned off, according to an agency recall notice.The company has made another software patch available to address the issue.

The device is intended to pump blood through an extracorporeal circuit for periods lasting less than six hours to provide either:

  • Full or partial cardiopulmonary bypass during open surgical procedures on the heart or great vessels; or
  • Temporary circulatory bypass for diversion of flow around a planned disruption of the circulatory pathway necessary for open surgical procedures on the aorta or vena cava.