# Medtronic Recalling Monitor Battery Pack
- URL: https://www.fdaweb.com/medtronic-recalling-monitor-battery-pack/
- Published: 2016-04-20T12:00:00.000Z
- Updated: 2026-09-14T20:56:03.000Z
- Author: David McFarland
- Tags: Devices, #legacy-id-D5135433

> Medtronic says it is recalling the battery pack in Covidien Oridion-labeled patient monitors. The company says that the recall is due to a battery manufacturing defect that may increase the risk of thermal damage in the battery pack. The packs were produced by a contract manufacturer between 4/2014 and 2/2016.  
>  
> Medtronic says it has received seven reports of thermal damage out of 9,817 battery packs impacted by the recall. Of these seven reports, it says, one involved a fire resulting in smoke inhalation and minor burns.  
>  
> Customers should use the monitors on AC power without the battery pack until a new battery pack is provided, the company says.