Medtronic Recalls Malfunctioning StealthStation Software
Medtronic is recalling (Class 1) its StealthStation S8 Application Versions 1.2.0, 1.1.0, 1.0.3, 1.0.2, and 1.0.1 after customer complaints about a software error that removed numbers or letters in displayed text and replaced them by a space, in either the “distance to target” or “tip stop point” display during navigation. “The missing digit may result in the mechanical depth stop being set too short or too long and may cause the surgeon to use an incorrect measurement for placement of the biopsy tip stop during navigation,” a recall notice says.
“Performing a cranial procedure using incorrect placement could lead to serious permanent injuries, including brain, nerve, or cerebrovascular damage, abnormal functions to the brain, spine, or muscles, paralysis, or death,” the notice says. To date, Medtronic reports 28 complaints related to the issue, which include no reported injuries or deaths.
The StealthStation is intended as an aid for locating anatomical structures during neurosurgery (treatment of diseases or disorders of the brain and spine), according to the company. It is indicated for use in any medical condition where stereotactic surgery of the brain or spine may be appropriate and where reference to a rigid anatomical structure, such as the skull, can be identified in relation to images of the anatomy, it adds.