# Medtronic Safety Notice on Imaging System
- URL: https://www.fdaweb.com/medtronic-safety-notice-on-imaging-system/
- Published: 2020-01-17T12:00:00.000Z
- Updated: 2026-09-14T15:59:06.000Z
- Author: David McFarland
- Tags: Devices, #legacy-id-D5145876

Medtronic has issued an urgent field [safety notice](https://www.bfarm.de/SharedDocs/Kundeninfos/EN/17/2020/15925-19%5Fkundeninfo%5Fen.pdf?%5F%5Fblob=publicationFile&v=1&ref=fdaweb.com) about the potential for navigational inaccuracy when its O-arm O2 Imaging System’s auto-registration feature is used in conjunction with an image-guided surgery system. “Medtronic has determined that replacement of the internal O-arm gantry tractor motor drive will resolve the issue of increased potential for loosening of the drive belt in the population of affected O-arm O2 Imaging Systems,” the notice says. Medtronic says it is scheduling service representatives to make necessary corrections.