More Hemodialysis Toxic Compound Analysis Needed: FDA

Share

FDA says that more testing and analysis are needed to determine if several toxic compounds are present in additional hemodialysis and peritoneal dialysis systems. Updating its 5/6/22 letter to healthcare providers on the evaluation of the potential risk of exposure to non-dioxin-like (NDL) polychlorinated biphenyl acids (PCBAs) and NDL polychlorinated biphenyls (BCPs) when using certain Fresenius Medical hemodialysis systems, FDA says it has become aware of other hemodialysis and peritoneal dialysis systems that have parts made of chlorinated peroxide cured silicone and it is unknown at this time whether there is a risk of exposure to NDL PCBAs or NDL PCBs.

The update says that in the Fresenius systems, the source of the toxic compounds was the silicone tubing used as part of the hydraulics in the dialysate lines. “Specifically,” it says, “the manufacturing process of the silicone tubing used a chlorinated peroxide initiator that resulted in NDL PCBAs and NDL PCBs.”

FDA says Fresenius modified the silicone tubing to platinum catalyst silicone tubing and the company reported that NDL PCBAs and NDL PCBs were not detected in machines with the modified tubing.

The notice says FDA is working with the additional manufacturers to further collect and evaluate data to determine the risk of exposure to NDL PCBAs and NDL PCBs in patients and develop strategies to address the issue if needed.