Tandem Updates Apple Mobile Diabetes Pump App
Tandem Diabetes Care says it is recalling version 2.7 of the Apple iOS t:connect mobile app used in conjunction with the t:slim X2 insulin pump with Control-IQ technology, and replacing it with app version 2.8.2. FDA says the recall is Class 1.
A company statement says the reason for the recall (correction) is due to an issue with the software that may cause the mobile app to crash and be automatically relaunched by the iOS operating system. “This cycle intermittently repeats, which leads to excessive Bluetooth communication that may result in pump battery drain and may lead to the pump shutting down sooner than typically expected,” the company says.
Tandem says there were 224 reported injuries and no deaths associated with the recall as of 4/15.