Vero Biotech Recalls Nitric Oxide System
Vero Biotech is recalling its Genosyl DS; Nitric Oxide Delivery System due to a software issue that leads to errors in gas delivery. “Typically, this issue caused delivery of lower-than-expected dosage of nitric oxide during the transition between primary and backup console,” a recall notice says. “If this happens, this could cause serious adverse events such as drops in oxygen level, heart problems, and clinical instability in the newborn.” The notice says there have been 11 complaints, three injuries and no deaths reported due to the issue.
The device is used to deliver a constant flow of nitric oxide, which is used to help open blood vessels and improve oxygen levels in critically ill newborns with respiratory failure. It is used with a ventilator in hospital or healthcare settings.