> ## Content Index
> Fetch the complete content index at: https://www.fdaweb.com/llms.txt
> Use this file to discover other available public pages before exploring further.

# Zoll Recalls Ventilators Over Operator Guides
- URL: https://www.fdaweb.com/zoll-recalls-ventilators-over-operator-guides/
- Published: 2024-06-21T12:00:00.000Z
- Updated: 2026-09-14T14:33:52.000Z
- Author: David McFarland
- Tags: Devices, #legacy-id-D5157230

Zoll Medical has recalled (Class 1) its Zoll 731 Ventilators (EMV+, AEV, Eagle II MRI compatible devices) and their operator and quick reference guides. A [recall notice](https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfRes/res.cfm?ID=207637&ref=fdaweb.com) says the guides omit patient safety ventilator MRI information, and they are being updated with instructions to ensure safe distance from the MRI for proper functioning of the ventilators and to reduce the risk of delay in ventilator therapy. It notes that the issue is under investigation by the firm.