Zyno Medical Recalls Infusion Pumps

Share

Zyno Medical has recalled certain Z-800, Z-800F, Z-800W and Z-800WF infusion pumps because they were released to customers with incorrect software versions that had not undergone required verification and validation testing. “As a result, these pumps can experience unexpected performance problems, including issues with essential functions and risk measures such as air in line detection, alarm volume, and reverse flow prevention,” a recall notice says. The worst-case scenarios of incorrect air in line detection and incorrect audio alarms associated with other failures could result in serious harms such as under- or over-delivery of drugs, and air in line that infuses a venous air embolism of up to 1mL, it says.

The pumps are used to provide intravenous infusion of parenteral fluids, blood, and blood products to a patient under the direction or supervision of a physician or other certified health care professional. To date, Zyno Medical has not reported any serious injuries or deaths, according to the recall notice.